Specification Sheet

Printable PDF :   AHP300 Specifications


Patient Range: Infant/Pediatric (5 kg) to Adult

Controls/Settings

Modes

  • (Pressure or Volume) A/CMV
  • (Pressure or Volume) SIMV
  • CPAP
  • Bi-Level Support (PS+CPAP) (can be programmed in custom mode)
  • NIV (non-invasive ventilation) available in all modes
  • 5 Programmable Quick Start Presets  

Tidal Volume:  40 to 2,000 ml

RR (Respiratory Rate): 0 or 5 to 60 b/min

I Time (Inspiratory Time): 0.5 to 2.0 sec

PEEP/CPAP: 0 to 25 cmH2O/0 to 25 mbar

PS (Pressure Support)**:  0 to 25 cmH2O / 0 to 25 mbar*

*PEEP + PS cannot exceed 25 cmH2O/mbar Operating Humidity 5 to 95% non-condensing

Flow: 5 to 60 L/min (flows > 80 LPM in PS Mode, CPAP and demand flows)

Pressure Control (Target Pressure): 15 to 55 cmH2O / 15 to 55 mbar

Ptrig (Sensitivity): -5 to -1 cmH2O / -5 to -1 mbar

Manual Inflation: 2 second maximum

FiO2: .21 to 1.00

Maximum I Time: 2 seconds

Flow Wave Pattern: Square or descending**

** Based on mode selected

Auto Lock Function: ON / OFF

Remote Alarm Option: Normally Open/Closed

Monitored Parameters

Paw (Peak, Base)
Delivered Tidal Volume
Spontaneous Breath Count
Integrated Battery Level
External Battery Level - When Attached

Calculated Parameters

I:E Ratio
Flow rate
FiO2

Alarms

High Airway Pressure: 10 to 55 cmH2O/10 to 55 mbar

Low Airway Pressure: 5 to 15 cmH2O/5 to 15 mbar

Back-up Ventilation Alert: 20 second fixed, apnea ventilation mode

Low Source O2Triggers at 40-37 psi

Low Source Air: Triggers at 40-37 psi

Device Alert: Malfunction alert

Critical Battery: Approx. 20 minutes of battery charge remains

Silence/Reset: 110 sec

Pneumatics

Gas Delivery System: Internal dual diaphragm

Air/Oxygen Mixer: Integrated

Oxygen Input: 41 to 87 psig oxygen

Power Requirements

AC Voltage: 110- 240 VAC, 50/60 Hz, max. 1A

DC Voltage: 11-15 VDC, max. 5A

Integrated Battery Parameters

Battery Composition: Sealed Lead Acid

Run time; 100% O27.5 hours

Run time; 60% O27.5 hours

Run time; 100% air (on compressor): 7 hours

W/ External Battery Pack Parameters (MCV-AUXBAT)

Battery Composition: Sealed Lead Acid

Run time; 100% O221 hours

Run time; 60% O221 hours

Run time; 100% air (on compressor): 21 hours

Dimensions & Weight

Width: 13.3" (34 cm)

Depth: 5.3" (13 cm)

Height: 13.8" (35 cm)

Weight: 18.8 lbs. (8.5 kg)

Environmental

Shock: 100 G-Force

Vibration: To IEC 60068-2-6

Operating Temperature: 0° F to 122° F (-18° C to 50° C)

Storage Temperature: -40° F to 140° F (-40° C to 60° C)

Operating Humidity: 5 to 95% non-condensing

Water Ingress: IPx 22

Signal Connections

Remote Alarm: 1/4" phone jack to connect to nurse call or remote system

Specialty Features

Quick Start:  Programmable presets to quickly recall pre-determined ventilation protocols

Custom Modes:  3 additional programmable custom modes that are available for quick recall

Oxygen Conserve:  Allows the use of up to 10 LPM of oxygen at 3 to 40 psi through the DISS port or up to 10 LPM of oxygen at less than 10 psi through the low pressure barb fitting for use with oxygen concentrator


Additional Language Downloads

Spanish

French

Russian

German